A large study to be published in the August issue of the American Journal of Medicine reports a record radiation dose reduction of 74.8% to MGH Cardiac CT patients during the past 6 years.

Evolution of Coronary Computed Tomography Radiation Dose Reduction at a Tertiary Referral Center

18/Jun/2012

Purpose
We aimed to assess the temporal change in radiation doses from coronary computed tomography angiography (CCTA) during a 6-year period. High CCTA radiation doses have been reduced by multiple technologies that, if used appropriately, can decrease exposures significantly.

Methods
A total of 1277 examinations performed from 2005 to 2010 were included. Univariate and multivariable regression analysis of patient- and scan-related variables was performed with estimated radiation dose as the main outcome measure.

Results
Median doses decreased by 74.8% (P<.001), from 13.1 millisieverts (mSv) (interquartile range 9.3-14.7) in period 1 to 3.3 mSv (1.8-6.7) in period 4. Factors associated with greatest dose reductions (P<.001) were all most frequently applied in period 4: axial-sequential acquisition (univariate: −8.0 mSv [−9.7 to −7.9]), high-pitch helical acquisition (univariate: −8.8 mSv [−9.3 to −7.9]), reduced tube voltage (100 vs 120 kV) (univariate: −6.4 mSv [−7.4 to −5.4]), and use of automatic exposure control (univariate: −5.3 mSv [−6.2 to −4.4]).

Conclusions
CCTA radiation doses were reduced 74.8% through increasing use of dose-saving measures and evolving scanner technology.

Source: The American Journal of Medicine